Sanyo MBR-704GR instruction manual Inner door Upper monitor bottle Lower monitor bottle

Page 9

REFRIGERATOR COMPONENTS

1.Front cover: Open this cover when connecting the remote alarm or replacing glow starter for fluorescent lamp.

2.Control panel: Panel can be opened when upper right corner of the lower part cover is pushed. Refer to page 10.

3.Automatic temperature recorder: A 7-day type recorder is provided. For proper usage of the recorder, see the instruction manual enclosed with the unit.

4.Access port: This port allows temperature measurement cables to enter the chamber from outside. A total of two (2) ports are provided; left side and right side.

5.Glass window: The window may have condensation in high humidity environment. Wipe off the condensation with a soft dry cloth.

6.Lock: Turn the key counterclockwise through 180o to securely lock the door.

7.Handle: Always hold the handle when opening/closing the door.

8.Fluorescent lamp: 20 W lamp. See page 19 for replacement.

9.Light switch: This switch is used for turning the fluorescent lamp off and on.

10.Air vent – front intake: Do not block this vent. If this vent is blocked, temperature regulation will become unstable. Do not insert a finger or any foreign object into this vent as there is danger from the internal fan.

11.Air vent – rear exhaust: Do not block this vent. If this vent is blocked, temperature regulation will become unstable. Do not place stored items in the path of the cold air.

12.Drawer (MBR-704GR only): Items to be stored in the chamber must be placed on the drawers. The maximum storage weight for each drawer is 40 kg.

Shelf (MBR-704G only): Items to be stored in the chamber must be placed on the shelves. The maximum storage weight for each shelf is 40 kg.

Do not put stored items directly onto the interior floor of the chamber. See page 14.

13.Leveling foot: Adjust the height of the leveling foot by turning the screw bolt until the unit is level.

14.Castors: When installing the unit the castors can be raised from the ground by using the leveling feet above.

15.Inner door

16.Upper monitor bottle

17.Lower monitor bottle

18.Door switch: The switch detects the door condition (open or close).

19.Condensation tray: The condensation on the door is accumulated in this tray. Wipe off the water occasionally.

20.Power switch (back side): Switch for the refrigerator. This switch also acts as an over-current breaker (10A).

21.Fixtures (back side): These keep the adequate space between the unit and wall and also can be used for fixing the unit. See page 12.

22.Evaporating tray (back side): Defrost water from the evaporator accumulates on the tray and evaporates into the atmosphere. See page 18 for cleaning.

23.Bottom air duct (MBR-704G only): The duct keeps temperature inside the chamber uniformly by making cooled air circulate. Put the duct at the bottom of the chamber. Do not put any articles under this duct. If this duct is blocked, temperature regulation will become unstable.

8

Image 9
Contents MBR-704G MBR-704GR Contents Precautions for Safe OperationSymbol shows Precautions for Safe Operation Prevent electric shock or injury May result Environmental Conditions Refrigerator Components Inner door Upper monitor bottle Lower monitor bottle Inside components of MBR-704G type Inside of front coverControl panel and keypad Installation site InstallationInstallation Fix the unitGround earth Preparation of monitor bottle Stock of containers START-UP of UnitExhaust air vent Load lineOperating Instructions Control panelDefrost cycles Temperature sensor calibration procedure Remote alarm terminal Alarms & Safety FunctionsAlarms and safety functions Alarm test procedureRoutine Maintenance Cleaning of cabinetCleaning of evaporation tray Replacement of fluorescent lamp Replacement of glow starterDisposal of Unit If nothing operates even when switched on Trouble ShootingAlarm device is activated When the unit does not get cold enoughPerformance SpecificationsSafety check sheet CEM