Sanyo MBR-704GR instruction manual Installation, Fix the unit, Ground earth

Page 13

INSTALLATION

Installation

1.Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a neutral detergent and wipe off with a soft cloth washed in clean water.

2.Adjust the leveling feet

Extend the leveling feet by rotating them as shown in the figure until they make contact with the floor. Ensure the unit is level.

Leveling foot

3. Fix the unit

Two fixtures are attached to the rear of the frame. Fix the frame to the wall by attaching a rope or chain between the wall and the fixtures.

If holes can be drilled in the wall, drill a 10.5mm diameter hole and fix the unit by using the special bolt-nut supplied. This bolt-nut can only be used on a concrete wall.

Bolt (provided)

45 mm

10.5 mm (diameter)

4. Ground (earth)

WARNING

Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is not grounded, it is necessary to install a ground by qualified engineers.

Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such

grounding may cause electric shock in the case of an incomplete circuit.

12

Image 13
Contents MBR-704G MBR-704GR Contents Precautions for Safe OperationSymbol shows Precautions for Safe Operation Prevent electric shock or injury May result Environmental Conditions Refrigerator Components Inner door Upper monitor bottle Lower monitor bottle Inside components of MBR-704G type Inside of front coverControl panel and keypad Installation site InstallationFix the unit InstallationGround earth Preparation of monitor bottle Stock of containers START-UP of UnitExhaust air vent Load lineControl panel Operating InstructionsDefrost cycles Temperature sensor calibration procedure Remote alarm terminal Alarms & Safety FunctionsAlarms and safety functions Alarm test procedureCleaning of cabinet Routine MaintenanceCleaning of evaporation tray Replacement of glow starter Replacement of fluorescent lampDisposal of Unit If nothing operates even when switched on Trouble ShootingAlarm device is activated When the unit does not get cold enoughPerformance SpecificationsSafety check sheet CEM