Sanyo MDF-593C, MDF-793C instruction manual Temperature Recorder

Page 25

TEMPERATURE RECORDER

8.Remove the sensor cover at the rear of the chamber wall. (Fig.

4)

9.Pass through the hole under the temperature control sensor. (Fig.

5)Tool for the hole: Screw driver (example) Hole depth: 135mm (5.3inch) Hole diameter: 13mm (0.5inch)

10.Route the recorder sensor into the chamber by sassing the hole. (Fig. 4)

11.Remove the sensor support and position the recorder sensor and then replace the stopper. (Fig. 4)

12.Seal around the hole for recorder sensor completely by applying the sealing compound (RTV) to outside and inside of the chamber. (Fig. 4)

13.Connect the power cord of the recorder to the terminal as shown in Fig. 6.

N

 

 

Neutral

 

L

 

 

Live

 

 

EEarth

14.Fix the grille with a screw and replace the side cover.

24

Fig. 4

Fig. 5

Fig. 6

Image 25
Contents MDF-593C MDF-793C Contents Symbol shows Precautions for Safe OperationMay cause bad effect on your health or environment Maintenance. These may be harmful to your health Water Precautions for Safe Operation Page Page Environmental Conditions Freezer Components Control panel cover LockControl panel Break down or power failure at the power supplyInstallation site InstallationAdjust the leveling foot InstallationRemove the packaging materials and tapes Fix the unitSTART-UP of Unit Operation after power failureRemote alarm terminal Temperature settingOperating Instructions Alarm temperature setting Setting of alarm resume time Alarm and safety function Alarm and safety functionsRough scheme Back-up systemFeeding of chart Temperature RecorderReplacement of cell Penholder Setting of back-up temperatureControl Panel Temperature Recorder Cleaning Routine MaintenanceCleaning of condenser filter Defrosting of inside wallTrouble Shooting Disposal of Unit Location of a nickel-cadmium battery Recycle of NICKEL-CADMIUM BatteryRemoval of nickel-cadmium battery Handling of batteryPerformance SpecificationsSafety check sheet CEM

MDF-593C, MDF-793C specifications

The Sanyo MDF-793C and MDF-593C are advanced laboratory refrigeration models designed for a variety of applications in research, medical, and industrial settings. Both units embrace cutting-edge technologies and engineering innovations to provide reliable performance while ensuring specimen preservation and optimal storage conditions.

One of the key features of the Sanyo MDF-793C is its ability to maintain a consistent temperature range from -10 to -30 degrees Celsius. This wide range allows users to store a variety of sensitive materials, including biological samples, medications, and vaccines. The unit is equipped with a microprocessor-controlled temperature management system, which offers precise temperature monitoring and control. Additionally, the unit includes a rapid cooling system that ensures quick recovery after door openings, minimizing temperature fluctuations that can compromise stored materials.

The MDF-593C, while slightly smaller in capacity, also offers impressive temperature control and stability. It is designed to have an internal temperature range of -10 to -25 degrees Celsius. Its compact design makes it suitable for labs with limited space, without compromising performance. Both models utilize environmental-friendly refrigerants, which align with current sustainability practices and regulations.

In terms of technological advancements, both models include digital displays that provide real-time temperature readings. These displays come with alarms that notify users of any aberrations, ensuring the highest levels of monitoring and security. The integration of a high-efficiency compressor not only supports rapid cooling but also contributes to lower energy consumption, making both the MDF-793C and MDF-593C cost-effective options for long-term use.

Another standout characteristic is the robust insulation used in the construction of both units. High-density polyurethane foam enhances thermal efficiency and minimizes energy loss. This feature is essential for maintaining optimal internal temperatures while reducing the overall operational costs associated with energy consumption.

Both models also feature adjustable shelving and user-friendly designs, allowing for easy organization and access to stored items. Their durable construction ensures reliability while a sleek external design allows them to fit seamlessly into various laboratory environments.

In summary, the Sanyo MDF-793C and MDF-593C are exemplary choices for those seeking efficient, reliable, and environmentally sustainable refrigeration solutions. Their advanced technologies, coupled with user-friendly features, make them essential tools in ensuring the integrity of valuable samples across various fields.