PRINTED CIRCUIT BOARD
1. CIRCUIT CHECK PROCEDURE
1.Low voltage transformer check
The low voltage transformer is located on the P.C.B. Measuring condition: Input voltage: 120V / Frequency: 60Hz
Terminal Voltage | LOAD | NO LOAD |
|
|
|
DC 12V | AC 25.8V | |
|
|
|
AC 3.4V | AC 4.0V | |
|
|
|
NOTE
1.Refer to Ciruit Diagram (point 4).
2.Secondary side voltage of the low voltage transformer changes in proportion to fluctuation of power source voltage.
3.The allowable tolerance of the secondary voltage is within ± 5% of nominal voltage.
2.Voltage Check
NO | CHECK POINT | REMARK |
1 | IC1 PIN 63, 64 | 5VDC |
2 | IC1 PIN 38 |
|
|
| 5V |
|
| 0V |
|
| T : 16.67ms(60Hz) |
3 | IC1 PIN 33 OR 34 |
|
|
| 5V |
|
| 0V |
|
| T : 250 ns(4MHz) |
- Key check point |
|
NO | MEASURE POINT | WAVE FORM | REMEDY | REMARK |
|
|
|
|
|
1 | MP1 | DC 5V±0.25V | Replace Q8, ZD3, EC2 | NO LOAD |
|
|
|
|
|
2 | MP2 | DC 12V±2.0V | Replace D12, 13, R25, EC5, EC4 | NO LOAD |
|
|
|
|
|
- Check method
NOTE
Each measure point must be measured with GND points.
34