Cornelius CH 1500-A Introduction, Unpacking and Inspection, Design Data, Chiller Installation

Page 3

GENERAL INFORMATION

INTRODUCTION

The REMCOR ”CH”Series Recirculating Liquid Chiller is designed to provide an accurate, reliable, and user-- friendly system for cooling a continuous flow of pure liquid and keep that liquid at a constant temperature in vari- ous closed loop or tank cooling applications.

The ”CH”Series Chiller consists of an air--cooled refrigeration system housed in a sturdy sheet metal frame and cabinet. A standard pump and insulated liquid reservoir package provides a complete liquid cooling and circu- lating system.

The ”CH”Series Chiller is designed to operate in a clean laboratory or industrial environment where ambient temperatures range from 40 to100°F (5 to 38°C). With proper installation, operation, and maintenance, the ”CH”Series Chiller will provide years of trouble free service.

UNPACKING AND INSPECTION

This unit was thoroughly inspected before leaving the factory and the carrier has accepted and signed for it. Any damage or irregularities should be noted at the time of delivery and immediately reported to the carrier. Re- quest a written inspection report from the Claims Inspector to substantiate any necessary claims. In the event that an immediate replacement is necessary, please contact REMCOR Chiller Sales at 1--800--551--4423.

DESIGN DATA

Table 1. Design Data

Cooling Capacity:

BTU/hr (W) at 80_F(27_C) and

70_F (21_C) Liquid Temperature.

CH951

12,000 (3,515)

CH1500

CH1502

 

 

18,000 (5,272)

18,000 (5,272)

 

 

 

 

CH1503

18,000 (5,272)

Compressor Horsepower

1 (.746 kW)

1 1/2 (1.12 kW)

1 1/2 (1.12 kW)

1 1/2 (1.12 kW)

Electrical Data: Voltage/Phase/Hertz/Amperage

230/1/60

11.0 Amps

230/1/60 13.0 Amps

230/3/60

16.4 Amps

460/3/60 10.0 Amps

Refrigerant Type:

R134a

R22

R22

R22

Reservoir Capacity

6.0gallons (22.7 6.0 gallons (22.7

liters)liters)

6.0gallons (22.7 liters)

6.0gallons (22.7 liters)

Physical Dimensions, Width ¢ Depth

¢Height:

22I¢ 26.5XI¢

38.25I (56cm ¢ 67cm

¢97cm)

22I¢ 26.5XI¢

38.25I (56cm ¢ 67cm

¢97cm)

22I¢ 26.5XI¢

38.25I (56cm ¢ 67cm

¢97cm)

22I¢ 26.5XI¢

38.25I (56cm ¢ 67cm

¢97cm)

Fittings:

Process Connections

3/4” FPT (S/S)

3/4IFPT (S/S)

3/4IFPT (S/S)

3/4IFPT (S/S)

CHILLER INSTALLATION

Location of Chiller

THE CHILLER MUST BE LOCATED NEAR A PROPERLY GROUNDED ELECTRICAL OUTLET. THE CIR- CUIT SHOULD BE FUSED AND NO OTHER ELECTRICAL APPLIANCE SHOULD BE CONNECTED TO THE CIRCUIT. ALL ELECTRICAL WIRING MUST CONFORM TO NATIONAL AND LOCAL ELECTRICAL CODES.

The Chiller must be located in a well ventilated, indoor area where ambient temperatures will remain above 40_F (5_C) and will never increase above 100_F (38_C). To obtain optimum cooling capacity, the ambient temperature should be at or below 80_F (27_C).

1

92266

Image 3
Contents Chiller CH Series Table of Contents Design Data Chiller InstallationIntroduction Unpacking and InspectionProcess Connections Electrical ConnectionsProcess Liquid Flow Phase Protection / 3--Phase Monitor OptionPhase Protection / 3--Phase Option Adjustment Chiller OperationCooling Operation Temperature Indicator/Controller ThermostatCondenser Chiller MaintenanceSTART--UP Thermostat ControlPump Motor Circulation SystemFilters/Strainers Troubleshooting 92266 Cabinet Section Exploded View Qty Name Cabinet SectionPart Qty Name Electrical BOX Assembly92266 Pump and Tank Assembly Warranty This page Left Blank Intentionally One Cornelius Place Anoka, Minnesota 612 800