Chiller Operation
WARNING: NEVER OPERATE THE CHILLER WITH THE PANELS REMOVED.
ALWAYS USE THE POWER SWITCH TO TURN OFF THE CHILLER WHEN IT IS NOT BEING
USED.
ALWAYS ENSURE THAT ALL AIR INLETS AND OUTLETS ARE FREE FROM OBSTRUCTION. BE SURE THAT THE RESERVOIR IS FILLED WITH FLUID PRIOR TO POWERING UP THE UNIT.
Process Liquid Flow
After ensuring that the system piping is free from obstruction, that all valves are open, and the reservoir is full, push the CONTROL POWER switch to the ”ON”position. The pump should now be operating.
On three phase units such as the CH1502 and CH1503, it is important to check the pump rotation. Remove the left side panel, push the POWER button in, and observe the motor shaft. Make sure that the shaft is rotating in the direction of the arrow indicated on the pump housing. If the rotation is incorrect, reverse two of the three in- coming power supply leads at the terminal strip.
NOTE: Running the pump in reverse for too long will result in permanent pump damage. The Phase
All Chillers with pumps are supplied with a pressure regulating valve on the pump discharge. This valve is pre- set at the factory to ensure that system pressure does not exceed the capabilities of the pump motor and/or pip- ing. If this valve requires adjustment, please contact the REMCOR Service Group for the proper setting procedure and pressures.
Once the flow has been established the thermostat can be programmed to the desired
Phase Protection / 3--Phase Monitor Option
The
This control is located in the enclosure labeled “Monitor”on top of the electrical box.
Phase Protection / 3--Phase Option Adjustment
The following procedure will allow the
Rotate the adjustment control fully clockwise, or until the red (Trip) indicator illuminated. Slowly rotate the adjustment control in a