Symmetricom 5401, 5402 manual Input Threshold High/Low

Page 29

Operating Manual

Operation

Description:

In Table 3 is a list of the data found in the response. The data is in the same format as the associated "Enter" command. A space exists between each of the parameters, except where noted. The total count is 176 characters.

Table 3 Response Format

Variable

Return String

Count

First

 

 

 

 

Dynamic Gain Floor

DFgg.g

6

1

 

 

 

 

Dynamic Gain Ceiling

DCgg.g

6

8

 

 

 

 

Dynamic Gain Rate Factor

DRrr.rrrr (in nanoseconds)

9

15

 

 

 

 

Dynamic Gain Offset Factor

DOoooo (in nanoseconds)

5

25

 

 

 

 

Km

KM.kkkkkk

9

31

 

 

 

 

Ko1

KO1.nnnnnn

10

41

 

 

 

 

Ko2

KO2.oooooo

10

52

 

 

 

 

Disgain

DGddddd

7

63

 

 

 

 

Resync Threshold

RTrrrrrr.rrr (in microseconds)

12

71

 

 

 

 

Resync Delay

RDdddd (in microseconds)

6

84

 

 

 

 

Rate Threshold

RArrrrrr.rrr (in microseconds)

12

91

 

 

 

 

Burst

BUx (x = 0 = Off, x =1 = On)

3

104

 

 

 

 

Default Calibrate Unit

CZrr.rrrr (in microseconds)

9

108

 

 

 

 

1PPS Offset

POyyyyyy.yyy (in microseconds)

12

118

 

 

 

 

Input Threshold

ITx (x = 0 = High, x = 1 = Low)

3

131

 

 

 

 

Programmable Delay Value

PDddd

5

135

 

 

 

 

Oscillator Control Range

ORsrr.rrrr (in microseconds)

10

141

 

 

 

 

Design Verification

DVx (x = 0 = Off, x = 1 = On)

3

152

 

 

 

 

Rate Error TC

RCnnnnnn

8

156

 

 

 

 

Offset Error TC

Ocoooooo (no space here)

8

165

 

 

 

 

 

CR LF

2

174

 

 

 

 

Input Threshold High/Low

Command Format:

PN 8500-0150, Rev. A, May 2005

21

Image 29 Contents
Airborne Disciplined Rubidium Oscillator Symmetricom, Inc Operating Manual Preface Limited WarrantyProprietary Notice Where to Find Answers to Product and Document QuestionsLimitation of Liability Conventions Used in This Guide Operating Manual Preface Typographical Conventions When text appears It means This wayPN 8500-0150, Rev. A, May Contents-1 Operating Manual ContentsOperating Manual SpecificationsOperating Manual Introduction Product OverviewSymmetricom 5402 Disciplined Rubidium Oscillator Symmetricom 5401 Airborne Disciplined Rubidium OscillatorSymmetricom 5401 Installation Procedure InstallationOperating Manual Installation Unpacking and Inspection5401 Adjustment Locations Symmetricom 5401 AdjustmentsSymmetricom 5402 Rack Mounting Procedure Symmetricom 5402 Initial Operation Symmetricom 5402 Installation ProcedureSymmetricom 5402 Adjustments Operating Manual Operation RS-232 Input/OutputStatus Request Operating Manual Operation DRO Commands Via RS-232Description Discipline OffDynamic Gain Control Discipline OnMinutes Enter KMEnter KO1, KO2 Enter Resync Threshold Enter DisgainEnter Rate Threshold Enter Resync DelayBurst Status Off Burst Status OnStatus will be output every second Turns burst status offRequest TI DRO Discipline Sequence FlowchartDTyyy Firmware Number PN 8500-0150, Rev. A, May Enter DAC ValueRequest Firmware Version Operating Manual Operation Revision Letter of Firmware Enter Start-up Parameters Request Start-up Parameters Response Format Description Calibrate UnitReturns the unit to a default state Default Values Set DefaultsEnter 1 PPS Offset not implemented Request Setup ParametersInput Threshold High/Low Enter Oscillator Control Range not implemented Enter Programmable Delay ValueDesign Verify On/Off Enter Estimation Filter Time Constant Cal StatusEnter Dynamic Gain Factors Input Specifications and Characteristics Input Specifications and CharacteristicsOperating Manual Specifications Output Specifications and Characteristics Output Specifications and CharacteristicsCoherence a 5401 Indicators Front Panel Indicators Front Panel5402 Indicators Front Panel DRO Internal Time Bases10 MHz Oven Oscillator Operating Manual Specifications DRO Internal Time BasesMHz Oven Oscillator Weight DimensionsApproximately 8 pounds Approximately 10 pounds PN 8500-0150, Rev. A, MayEMI Environment