Care and Maintenance

Care and Maintenance

The 3M Electronic Pipettor - 5mL requires regular cleaning to ensure trouble-free operation. Use a soft cloth lightly moistened with a mild detergent to clean the outer surface of the pipettor. DO NOT AUTOCLAVE.

Cleaning the Tip Cone

Monthly tip cone cleaning is recommended. However, with heavy use, clean more frequently. If liquid enters the shaft, clean tip cone immediately.

1.Switch the unit “OFF”.

2.Remove the tip ejector collar: Hold down the tip ejector lever and twist off the tip ejector collar.

3.Use a soft, lint-free wipe lightly moistened with ethanol to gently clean the tip cone.

4.Replace the tip ejector collar allowing the ejector handle connection to snap into the attachment notch of the ejector collar.

tip cone

filter

ejector collar

Page 12 of 15 - June 2001

Care and Maintenance

Filter Replacement

If the pipettor is not performing due to a clogged filter or if sample fluid has contaminated the filter, the filter should be replaced.

1.Remove the contaminated or clogged filter by simply pulling it straight out of the end of the tip cone (see Figure 1).

2.Clean the outside surface and the inner filter placement surface of the tip cone using a swab moistened with ethanol (see Figure 2).

Removing Contaminated or

Cleaning the Tip Cone Before

Clogged Filter from Tip Cone

Installing New Filter

Figure 1

Figure 2

3.Place the pipettor in the charging stand for 15 minutes to allow drying of the tip cone.

4.Place a new filter into the tip cone (see Figure 3).

Figure 3