Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document.www.spotontemperature.com reorder #202052A
6 | Introduction |
3.To reduce the risks associated with impact and facility medical device damage:
•Do not use the SpotOn control unit as a handle to transport or to move the device to which it is attached.
•Do not initiate temperature monitoring unless the SpotOn control unit is safely placed on a hard, flat surface or is securely mounted.
4.To reduce the risks associated with environmental contamination:•Follow applicable regulations when disposing of this device or any of its electronic components.
NOTICES1.The SpotOn temperature monitoring system meets medical electronic interference requirements. If radio frequency interference with other equipment should occur, connect the unit to a different power source.
2.Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional.
3.To avoid SpotOn temperature monitoring system damage that may impact performance:
•Do not store the SpotOn control unit and system components in a wet or damp place.
•Do not spray cleaning solutions onto the control unit or into the sensor cable connector.
•Do not immerse the SpotOn temperature monitoring control unit or system components in any liquid or subject them to any sterilization process.
•Do not use solvents such as acetone or thinner to clean the control unit; avoid abrasive cleaners.
•Clean control unit exterior with a damp soft cloth and 70% mixture of isopropyl alcohol and water. For full cleaning procedures see “Section 5: General Maintenance, Storage and Service” on page 17.
•Do not immerse any of the devices or use a dripping wet cloth for cleaning.4.The SpotOn temperature monitoring sensor is not made with natural latex material.
5.To the full extent permitted by law, the manufacturer and/or importer declines all responsibility for injury resulting from the unit being used in conjunction with unapproved system components.
Proper use and maintenance
Arizant Healthcare Inc., a 3M company assumes no responsibility for the reliability, performance, or safety of the temperature monitoring system if the following events occur:
•Modifications or repairs are performed by unqualified personnel.•The unit is used in a manner other than that described in the Operator’s Manual.
•The unit is installed in an environment that does not meet the appropriate electrical requirements.
Read before servicing equipment
•All repair, calibration, and servicing of this equipment must be performed by 3M Patient Warming.
•There are no user serviceable parts inside the equipment or power supply.English 202052A