Electrical characteristics

External power supply100-240 VAC, 50-60 Hz Output; 5 VDCClass II, double insulated medical gradeMaximum heating power2 W

Temperature characteristics

Over-temperature cutoff43°C - Skin thermistor reading44.5°C - Heater thermistor reading

Performance characteristics

Equilibration time3 minutesTime responseApproximately three minutesMeasurement locationsPatient’s lateral forehead above the orbital ridgeMeasurement range25°C - 43°CMeasurement readoutCelsius or Fahrenheit

Leakage current

CF classification. Meets leakage current requirements in accordance with UL 60601-1 and EN60601-1.

Specifications

Environmental conditions

Ambient temperature range10°C to 40°C (50°F to 104°F)Storage and transport temperature range-20°C to 60°C (-4°F to 140°F)Store all components at room temperature and in a dry place when not in use.Operating humidity10 to 75% RH, noncondensingAtmospheric pressure range80 kPa to 106 kPa (12 PSI to 15 PSI)Oxygen environment

The sensor only is suitable for use in the presence of an oxygen enriched environment.

English 202052A