16Troubleshooting

Errors

Error Screen

Condition

Cause

Action

 

 

 

 

 

E-1 through E-16, E-18

Sensor not properly

Make sure the sensor is connected to the

 

 

 

 

 

 

 

 

 

 

through E-99

connected to the sensor

sensor cable.

 

 

 

 

 

 

cable.

 

 

E-xx

 

 

Sensor failure.

Replace sensor.

 

 

 

 

 

 

Cable failure.

Replace sensor cable.

 

 

 

 

 

 

 

Contact 3M Patient Warming technical

 

 

 

 

 

 

 

service.

 

 

 

 

 

E-17

Sensor failure.

Replace sensor.

 

 

 

 

 

 

 

 

 

 

 

Cable failure.

Replace sensor cable.

 

 

 

 

 

 

 

Contact 3M Patient Warming technical

 

 

 

 

 

 

 

 

E-xx

 

 

 

service.

 

 

 

 

 

 

Control unit failure.

Discontinue use of the unit. Contact a

 

 

 

 

 

 

 

biomedical technician.

 

 

 

 

 

E-100 and up

Control unit failure.

Discontinue use of the unit. Contact a

 

 

 

 

 

 

 

 

 

 

 

 

biomedical technician.

 

E-xx

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 