Cleaning and Disinfecting Procedure

CAUTION:

Do not immerse any of the devices or use a dripping wet cloth for cleaning. Moisture may seep inside the device and damage the electrical components and lead to incorrect temperature reporting.

Do not spray cleaning solutions onto the control unit or into the sensor cable connector. Damage to the control unit or sensor cable connector may occur.

Cleaning the control unit and cable

Clean the SpotOn control unit on an as-needed basis or per facility policies and procedures for cleaning electronic equipment. Clean the sensor cable between each use.

1.Disconnect power supply from power outlet.

2.Use a slightly damp soft cloth moistened with a mild, nonabrasive cleaning solution to clean the device surfaces, cords, and cables. Avoid getting liquid into electronic ports.

3.Dry with a separate soft cloth.
Disinfecting the sensor cable
1.Clean the devices as described above.

2.Wipe down the control unit and cable using a damp soft cloth and 70% isopropyl alcohol solution. Avoid getting liquid into electronic ports.

3.Dry with a separate soft cloth.

Storage

Notice: Do not store the SpotOn control unit and system components in a wet or damp place. Damage to the electrical components may occur.

Store all SpotOn components at room temperature and in a dry place when not in use.

Service

All service must be performed by 3M Patient Warming or an authorized service technician. Call 3M Patient Warming Customer Service at 1-800-733-7775 for service information. Testing of the SpotOn system can be performed by 3M Patient Warming or call Customer Service for information about specific test procedures.