Check the 3MTM SpotOnTM Temperature Monitoring System website to ensure you have the most recent version of this document. | www.spotontemperature.com reorder #202052A |
| Troubleshooting | 15 |
Section 4: Troubleshooting
On Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Displays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
The following conditions are listed in the order of which troubleshooting actions should be performed.
On Mode
| Condition | Cause | Action |
Unit does not power up. | Unit is not plugged in, or power supply | Make sure the power supply is plugged into |
| is not plugged into an appropriate | the control unit. Make sure the control unit |
| outlet. | is plugged into an appropriate power outlet. |
| Unit failure. | Contact 3M Patient Warming technical |
|
| service. |
Displays
Condition | Cause | Action | ||||
“Ready” screen continues to display | Sensor not properly adhered to the | Make sure the sensor is connected to the | ||||
while sensor is adhered to the patient. | patient. | patient. | ||||
|
|
|
|
| Low patient temperature | Press and hold the °C/°F button on the back |
|
|
|
|
| (below 29°C). | of the control unit for five seconds to force |
|
|
|
|
|
| the control unit into equilibration mode. The |
|
|
|
|
|
| |
|
|
|
|
|
| patient’s temperature will then display. |
| SN - 000000000 |
| Control unit failure. | Discontinue use of unit. Contact 3M Patient | ||
|
|
|
|
|
| Warming technical service. |
“Running” screen does not appear on | Control unit may still be equilibrating. | Wait until the SpotOn control unit has | ||||
the control unit. |
| finished equilibration. Continue use. | ||||
|
|
|
|
| Control unit failure. | Contact 3M Patient Warming technical |
|
|
|
|
|
| service. |
|
|
|
|
|
|
|
Patient temperature does not appear on the patient monitor.
36 9 C36 0 C39
37 
35 
Control unit may still be equilibrating. Wait until the SpotOn control unit has finished equilibration. Continue use.
Monitor cable is not plugged in. | Make sure the monitor cable is plugged into |
| the control unit and patient monitor. |
|
|
Cable failure. | • Replace monitor cable. |
| • Contact 3M Patient Warming technical |
| service. |
Control unit failure. | Contact 3M Patient Warming technical |
| service after use. |
The patient’s temperature output on | Cable failure. | • | Replace monitor cable. |
the control unit and patient monitor |
| • | Contact 3M Patient Warming technical |
does not read the same value. |
|
| service. |
English 202052A