Bair Hugger Model 505 Temperature Management Unit Operator’s Manual

≤44 in. (112 cm)
Safety strap
≥28 in. (71 cm)
Figure E. Model 505 unit attached to an IV pole
Figure F. Model 505 unit attached to a bedrail

Mounting the Model 505 Temperature Management Unit

using an iv pole

The Model 505 unit clamps easily to an IV pole (see Figure E). Simply turn the handle clockwise to tighten the clamp onto an IV pole, counterclockwise to release.

Warning: To prevent tipping, clamp the Model 505 temperature management unit to an IV pole at a height that provides stability. We recommend clamping the unit no higher than 44 in. (112 cm) from the floor on an IV pole with a minimum 28 in. (71 cm) diameter wheelbase. Failure to do so may result in IV pole tipping, catheter site trauma, and patient injury.

using a bedrail

The Model 505 unit can also hang on the edge of a bed. The safety strap is designed to loop around the bedrail, keeping the Model 505 unit safely suspended even if the unit is inadvertently dislodged from the bedrail (see Figure F).