?WHAT DISPLAY SYMBOLS MEAN

Display/

Symbol

Battery Full

Battery Low

PULSE DISPLAY ERROR

Condition/Cause

The symbol shown while measure- ment is in progress. It blinks while detecting the pulse.

An irregular heartbeat or body movement may have occurred. Refer to page E-9 for more infor- mation on irregular heartbeats.

The battery power indicator is displayed during measurement.

The batteries are low when it blinks.

Unstable blood pressure due to excessive body movement.

The systolic and diastolic values are within 10 mmHg of each other.

The pressure value did not increase during inflation.

The cuff is not fastened correctly.

There is an air leak in the cuff or monitor.

The pulse is not detected correctly.

Cuff Inflation Meter

Recommended

Action

Measurement is in progress, remain as still as possible.

Take measurement again and consult with your physician.

Replace all batteries with new ones when the symbol blinks. Alkaline batteries recommended.

Try the measurement again. Remain very still during the measurement.

Fasten the cuff correctly, and try the measurement again.

Check for air leaks along the tube and around the air socket.

Refasten the cuff and retake measurement.

Make sure tube is properly connected to cuff and monitor.

Try the measurement again. Remain very still during the measurement.

Measurement is in progress, remain as still as possible.