7.0 USER PARAMETERS
The scale has 7 User Parameters that can be set by the user. To set
these Parameters press the [Func] key. The display will show the
first parameter, "F0 H-L".
Pressing the [Tare/Ï] will cycle through the other parameters.
Pressing [Zero/Enter] will allow you to enter a parameter directly for
setting or further enter a sub-parameter. Press the [Print/Esc] key to
leave the parameter unchanged.
For setting a new value for the parameter or a sub-parameter, first
enter the parameter or sub-parameter by pressing the [Zero/Enter]
key. Then use the [Pcs/Í] key or the [Units/Î ] key to move the
active/flashing digit to the left or to the right and use the [Tare/Ï]
key to increment the flashing digit. Press the [Zero/Enter] key to
enter the changed value.
Use the [Print/Esc] key to leave the parameter.
For example,
When the display shows “F0 H-L”, press the [Zero/Enter] key.
The display will show “SEt Lo” (the sub-parameter). Press
[Zero/Enter] to set the low limit or press the [Tare/Ï ] to skip to “SEt
Hi” for setting the high limit. When all digits have been set, press
[Zero/Enter] to store the value. Display will go back to the sub-
parameter just set, i.e. “SEt Lo” or “SEt Hi”.
Press [Print/Esc] to return to “FO H-L”.
Advance to the next parameter (F1 toL) by pressing [Tare/Ï] if
needed or press [Print/Esc] to return to weighing.