© Adam Equipment Company 2012 Page 8 Rev C5, January 2012

5.2 MEASURING HEIGHT

While measuring the height, it is necessary to pull up the measuring
board at a right angle with the inside tube.
When the tube is pulled out straight, it is sufficient to measure the
height form 80-136cm. The number can b e obtained upon the conjoint
place where the upper part of middle tube screw meets with the inside
tube scale.
Further, if the middle tube is p ulled out straight, it is possible to
measure the height from 136-210cm. The reading can be obtained at
the conjoint place where the upper part of outside tube screw meets
with the middle tube scale.
6.0 CALIBRATION
Before calibrating the scale, you should ensure that you have a known weight for
calibration.
1. When in normal weighing mode with the scale at zero press and hold down
[TARE] and [ON / OFF] keys to enter the calibration mode.
2. If the calibration switch is in the off position on the main PCB inside the scale, the
indicator will show ‘CAL.OFF” and then exit this mode. If the indicato r shows”
CAL-?”, the scale is ready for calibration, showing th e calibration switch is in the
on position. If the calibration switch is in the Off p osition open up the indicator
and move the switch to the On position. See diagram on page 8.
3. When the indicator shows ” CAL-?”, press the [TARE] key to confirm and go to
next step, or press the [ON / OFF] key to exit the calibration mode.
4. When 0.0’ is displayed the scale will begin to calibrate the scale’s zero-point.
Ensure that there is no load or weight on the scale’s platform . Press the [TARE]
key to confirm, or press the [ON / OFF] key to exit this mode.
5. A few seconds after the [TARE] key has been pressed in step 4 the scale will