![](/images/new-backgrounds/1148274/14827439x1.webp)
Chapter 7O Agilent 10737L and Agilent 10737R Compact
Interferometers
Procedure
Figure 7O-9. Agilent 10737L Compact Three-Axis Interferometer — return beam pattern
12Install the receiver assembly.
To do this, reverse the “Removing the receiver assembly” procedure, above.
13Plug in the
14Adjust each receiver’s gain by turning its gain adjustment screw to cause the receiver’s LED to light, then reduce the gain until the LED just turns off. For more information, see Agilent 10780F instructions in Chapter 8, “Receivers,” of this manual.