Upgrading an HP 1100 Series LC

The Xcalibur 2.0 and higher data system supports control of the Hewlett-Packard (HP) 1100 Series LC only if you upgrade the HP 1100 communication interface to an Ethernet interface.

YTo upgrade your HP 1100 Series LC

1.Order the appropriate upgrade parts from your local office for Thermo Electron San Jose products:

If the HP 1100 is not interfaced to an MS detector as an Xcalibur-controlled GPIB inlet device, order the JetDirect Ethernet Control kit, which contains the parts listed in Table 4.

If the HP 1100 is interfaced to an MS detector as an Xcalibur-controlled GPIB inlet device, order the parts listed in Table 5.

Note An HP 1100 LC system that is currently interfaced to a MS detector as an Xcalibur-controlled GPIB inlet device should have already installed the contact closure PCB and the external contact closure cable. Therefore, you need to order only the parts specified in Table 5.

Table 4. JetDirect Ethernet Control Kit (P/N OPTON 30018)

Part Number

Description of Kit

 

 

00012-27714

Contact closure PCB

 

 

00012-27716

External contact closure cable

 

 

00012-70008

Ethernet 10 Base-T cable (2)

 

 

00825-01-00024

10/100 Autosensing 8-port Ethernet switch

 

 

00825-01140

HP JetDirect 400N PCB

 

 

Thermo Scientific

Agilent 1100 Series LC Devices Connecting the Hardware and Triggering Data Acquisition