Performance Characteristics (continued)

Vertical System: Analog Channels (continued)

Range1,2

54621A/D, 54622A/D, 54624A: 1 mV/div to 5 V/div

 

54641A/D, 54642A/D: 2 mV/div to 5 V/div

 

 

Maximum Input

CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk

 

With 10073C/10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk

 

5 Vrms with 50 Ω input

 

 

Offset Range

54621A/D, 54622A/D, 54624A:

 

±5 V on ranges <10 mV/div; ±25 V on ranges 10 mV/div to 199 mV/div; ±100 V on ranges ≥200 mV/div

 

54641A/D, 54642A/D:

 

±5 V on ranges <10 mV/div; ±20 V on ranges 10 mV/div to 200 mV/div; ±75 V on ranges >200 mV/div

 

 

Dynamic Range

Lesser of ±8 div or ±32 V from center screen

 

 

Input Resistance

54621A/D, 54622A/D, 54624A: 1 MΩ ±1%

 

54641A/D, 54642A/D: 1 MΩ ±1%, 50 Ω selectable

 

 

Input Capacitance

54621A/D, 54622A/D, 54624A: ~ 14 pF

 

54641A/D, 54642A/D: ~ 13 pF

 

 

Coupling

54621A/D, 54622A/D, 54624A: ac, dc, ground

 

54641A/D, 54642A/D: ac, dc

 

 

BW Limit

54621A/D, 54622A/D, 54624A: ~ 20 MHz selectable

 

54641A/D, 54642A/D: ~ 25 MHz selectable

 

 

Channel-to-Channel Isolation

54621A/D, 54622A/D, 54624A: dc to 20 MHz > 40 dB; 20 MHz to max bandwidth > 30 dB

(with channels at same V/div)

54641A/D, 54642A/D:DC to max bandwidth >40 dB

 

 

Probes

54621A/D, 54622A/D, 54624A: 10:1 10074C shipped standard for each analog channel

 

54641A/D, 54642A/D: 10:1 10073C shipped standard for each analog channel

 

 

Probe ID (Agilent/HP and

Auto probe sense

Tek compatible)

 

 

 

ESD Tolerance

±2 kV

 

 

Noise Peak-to-Peak

54621A/D, 54622A/D, 54624A: 2% full scale or 1 mV, whichever is greater

 

54641A/D, 54642A/D: 3% full scale or 3 mV, whichever is greater

 

 

Common Mode Rejection Ratio

20 dB @ 50 MHz

 

 

DC Vertical Gain Accuracy*1,2

±2.0% full scale

DC Vertical Offset Accuracy

54621A/D, 54622A/D, 54624A:

 

<200 mV/div: ±0.1 div ±1.0 mV ±0.5% offset value; ≥200 mV/div: ±0.1 div ±1.0 mV ±1.5% offset value

 

54641A/D, 54642A/D:

 

≤200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value; >200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value

 

 

Single Cursor Accuracy1,2

±{DC Vertical Gain Accuracy + DC Vertical Offset Accuracy + 0.2% full scale (~1/2 LSB) }

 

54621A/D, 54622A/D, 54624A example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,

 

accuracy = ±{2.0%(80mV) + 0.1 (10 mV) + 1.0 mV + 0.5% (5 mV) + 0.2%(80 mV)} = ± 3.78 mV

 

 

Dual Cursor Accuracy*1,2

±{DC Vertical Gain Accuracy + 0.4% full scale (~1 LSB)}

 

Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,

 

accuracy = ±{2.0%(80 mV) + 0.4%(80 mV)} = ±1.92 mV

 

 

* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

1For 54621A, 54621D, 54622A, 54622D, and 54624A, 1 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting.

2For 54641A, 54641D, 54642A, and 54642D, 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 2 mV/div sensitivity setting.

14

Page 14
Image 14
Agilent Technologies 54600-Series manual 54641A/D, 54642A/DDC to max bandwidth 40 dB