4.Change the class label for S11A: LABEL CLASS, S11A, ERASE TITLE.

5.Enter the label of PSHORT 1 by using the knob, the SELECT soft key and the SPACE soft key.

6.Complete the label entry procedure: TITLE DONE, LABEL DONE.

Follow a similar procedure to enter the remaining standard classes and labels as shown in the table below:

Finally, change the cal kit label as follows:

1.Press LABEL KIT, ERASE TITLE.

2.Enter the title “P BAND.”

3.Press TITLE DONE, KIT DONE (MODIFIED). The message “CAL KIT SAVED” should appear.

This completes the entire cal kit modification for front panel entry. An example of programmed modification over the GPIB bus through an exter- nal controller is shown in the Introduction To Programming section of the Operating and Service manual (Section III).

Standard

Standard

Class

class

numbers

label

 

 

 

S11B

2

PSHORT 2

S11C

3

PLOAD

S22A

1

PSHORT 1

S22B

2

PSHORT 2

S22C

3

PLOAD

FWD TRANS

4

THRU

FWD MATCH

4

THRU

REV TRANS

4

THRU

REV MATCH

4

THRU

RESPONSE

1,2,4

RESPONSE

 

 

 

25

Page 25
Image 25
Agilent Technologies 8510-58 manual Pshort