Image 15

In full rate sample mode with the use of appropriate lowpass filtering (Table

3)signal aliasing is completely avoided. In this case the “Operating Region” is renamed “Alias Exposure Free Zone” to indicate this special condition. NOTE: Does not apply to HF operation.

Although the operating regions for all digitizer models are similar, each model or model family may have important differences in their operating region curves due to differences in maximum sample rate, front-end analog performance, and memory option. Appendix C provides the operating curves for all supported digi- tizers and most memory options.

Operating region: with span and memory constraints

 

1.E+01

 

 

 

 

 

 

 

 

 

 

 

1.E+00

 

 

 

 

 

 

 

 

 

 

 

1.E-01

 

 

 

 

 

 

 

 

 

 

(sec)

1.E-02

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Time record length

1.E-03

 

 

 

 

 

 

 

 

 

 

1.E-04

 

 

 

 

 

 

409601

pointsfrequency of

 

 

 

 

 

Operating region

204801

 

 

 

 

 

 

102401

 

 

 

 

 

 

 

 

51201

 

1.E-05

 

 

 

 

 

 

25601

 

 

 

 

 

 

 

12801

 

 

 

 

 

 

 

 

6401

 

 

1.E-06

 

 

 

 

 

 

3201

 

 

 

 

 

 

 

 

 

1601

 

Number

 

 

 

 

 

 

 

 

 

801

 

 

 

1.E-07

 

 

 

 

 

 

401

 

 

 

 

 

 

 

 

 

201

 

 

 

1.E-08

 

 

 

 

 

 

Minimum

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.E+02

1.E+03

1.E+04

1.E+05

1.E+06

1.E+07

1.E+08

1.E+09

1.E+10

1.E+11

Span (Hz)

Figure 6. The true operating region is fully described when constrained by all the previous parameters: number of frequency points, span, and digitizer memory. See Appendix C for operating region charts for the various supported digitizers.