Current Limit Characterization

Current limit measurements demonstrate the degree to which a constant voltage power supply limits its maximum output current to a preset value. This preset value can be fixed or variable throughout a specified range. There are basically three types of current limiting design implementations:

1.Conventional current limiting power supplies

2.CV/CC mode power supplies

3.Foldback current limiting power supplies

Conventional current limiting power supplies and CV/CC mode power supplies are very similar in function. These implementations generally vary only in the degree of regulation in the constant current operating region (see Figure 5) and in the ability of the user to adjust the CC operating point (CV/CC power supplies). A rounded crossover knee and sloping current limit characteristic denotes less precise current regulation. In comparison, a sharp knee and vertical current limit characteristic denote a higher degree of current regulation. The foldback current limiting power supply employs a technique that enables both the output voltage and current to decrease simultaneously for load resist- ances below the crossover value. The purpose of current limiting is to provide protection for the power supply and the device being powered (assum- ing the current limit value is below the maximum current rating of the device).

Test Overview/ Procedures

A measurement of the output voltage and current of the power supply under test is required while decrementing the electronic load resistance (or current in CC mode) by steps from an initial value that produces the power supply’s full rated voltage output (see Figure 6). The voltage will remain con- stant until the compliance current (output current of the power supply) increases to the preset current limit value. The crossover region or current limit has been reached when the rated output voltage of the power supply changes by a degree greater than the load regulation specification. At the current limit knee, the compliance current and output volt- age behavior is determined by the type of current limiting circuit implemented in the power supply design (see Table 3).

Table 3. Typical Test Results of Standard Current Limiting Implementations

 

I Compliance (or Iout)

Current Limiting Method

at Minimum Load Resistance

 

 

CV/CC

Remain constant (CC mode)

Conventional Current Limiting

Typically (105%) Imax

Current Foldback

Typically foldback is (50%) Imax

 

 

Figure 5. Typical Operating Characteristics of Three Types of Current

 

Limiting Power Supplies

7

Page 7
Image 7
Agilent Technologies AN 372-1 specifications Current Limit Characterization, Cv/Cc

AN 372-1 specifications

Agilent Technologies AN 372-1 is a versatile and highly regarded instrument in the field of electronic testing and measurement. Designed primarily for research and development, as well as validation in production environments, this advanced technology showcases Agilent's commitment to providing high-quality measurement solutions.

One of the standout features of the AN 372-1 is its wide frequency range capability, enabling it to effectively measure signals across various applications. This instrument supports a frequency range extending up to 26.5 GHz, making it suitable for high-frequency applications in telecommunications, aerospace, and defense sectors. With this broad frequency capability, users can rely on the AN 372-1 for accurate and reliable measurements in the most demanding environments.

Another significant attribute of the AN 372-1 is its high dynamic range, which allows for precise detection of both weak and strong signals in the presence of noise. This characteristic is essential for engineers working to ensure signal integrity in complex systems, as it allows for the identification of issues that may otherwise go unnoticed. Coupled with a fast sweep speed, the device can quickly capture transient events or changes in signals, making it a vital tool for time-sensitive applications.

The user interface of the AN 372-1 is designed for ease of use, with a touchscreen display that simplifies interaction and data visualization. Its intuitive design enables users to set up measurements and navigate through the many options effortlessly. Furthermore, the instrument is equipped with advanced analysis features and software that enhance data interpretation and manipulation, allowing engineers to derive meaningful insights from their measurements quickly.

In addition to its measurement capabilities, the AN 372-1 supports advanced connectivity options, including USB, LAN, and GPIB. This ensures seamless integration with other instruments and systems, facilitating a more streamlined testing workflow. The ability to connect with external software also enhances its functionality, allowing users to automate processes and improve test efficiency.

The AN 372-1's rugged design ensures it can withstand the challenges of both laboratory and field use. Its build quality speaks to Agilent's commitment to creating long-lasting instruments that offer reliable performance throughout their lifecycle. Overall, the Agilent Technologies AN 372-1 stands out as a comprehensive tool for engineers and researchers requiring precise, reliable, and efficient electronic measurements across a myriad of applications.