Regulatory Information

Installation Guide

Regulatory Information

Regulatory Information

Radio-Frequency Electromagnetic Field Immunity - IEC 61000-4-3: 1995 passes criterion B

According to IEC 61000-4-3: 1995, degradation of system noise perfor- mance may be observed in some narrow frequency ranges when a 3 V/m radio-frequency electromagnetic field is applied to the 86100A,B mainframe with an 86102U, 86106B, or 86107A module installed.

Radio-Frequency Electromagnetic Field Immunity - IEC 61000-4-3: 1995 passes criterion C

According to IEC 61000-4-3: 1995, the degradation of normal performance within the specification limits may be observed in the 86100A,B mainframe with an 86107A module installed. The instrument software recognizes this temporary signal degradation and warns the user that the data is suspect.

The warning prompts the user to press the Clear Display button. The user can then resume data acquisition. This required operator intervention results in a criterion C classification.