Voltage and Current Values
|
| Table |
| |||
Agilent Model | Max. Prog. |
|
| Max. Prog. | ||
| Voltage | Voltage | Current |
|
| Current |
E4350B | 60V | 61.5V | 8A |
|
| 8.16A |
E4351B | 120V | 123V | 4A |
|
| 4.08A |
Max. Prog.
Overvoltage
73.0V
140V
Transient Recovery Time
This test does not need to be performed on the Agilent E4350B.
CC RMS Noise Measurement Test Setup
IP VP IM IM I P S S
- - | +240 VDC MAX | - |
| I | SET TO |
+ | + - | + - | |||
|
|
|
| LOCAL | REMOTE |
|
|
|
| REMOTE |
|
MONITOR RESISTOR 1.6 OHM (Adjusted to 1 Ohm +/- 1%)
LOAD RESISTOR
E4350B 2 16 Ohm resistors in parallel (Adjusted to 7 Ohms +/- 1%)
E4351B 2 16 Ohm resistors in series (Adjusted to 25 Ohms +/- 1%)
RMS
VOLTMETER
Figure 2-3. CC RMS Noise Measurement Test Setup
Service Addendum | 5 |