Agilent Technologies E4350B Voltage and Current Values, Transient Recovery Time, Rms Voltmeter

Models: E4351B E4350B

1 57
Download 57 pages 1001 b
Page 5
Image 5
Voltage and Current Values

Voltage and Current Values

 

 

Table 2-3. Voltage and Current Values

 

Agilent Model

Full-scale

Max. Prog.

Full-scale

 

 

Max. Prog.

 

Voltage

Voltage

Current

 

 

Current

E4350B

60V

61.5V

8A

 

 

8.16A

E4351B

120V

123V

4A

 

 

4.08A

Max. Prog.

Overvoltage

73.0V

140V

Transient Recovery Time

This test does not need to be performed on the Agilent E4350B.

CC RMS Noise Measurement Test Setup

IP VP IM IM I P S S

- -

+240 VDC MAX

-

 

I

SET TO

+

+ -

+ -

 

 

 

 

LOCAL

REMOTE

 

 

 

 

REMOTE

 

MONITOR RESISTOR 1.6 OHM (Adjusted to 1 Ohm +/- 1%)

LOAD RESISTOR

E4350B 2 16 Ohm resistors in parallel (Adjusted to 7 Ohms +/- 1%)

E4351B 2 16 Ohm resistors in series (Adjusted to 25 Ohms +/- 1%)

RMS

VOLTMETER

Figure 2-3. CC RMS Noise Measurement Test Setup

Service Addendum

5

Page 5
Image 5
Agilent Technologies E4350B Voltage and Current Values, Transient Recovery Time, CC RMS Noise Measurement Test Setup