Procedure

Option Installation

Procedure

Saving System calibration Data from Removable Hard Disk to USB Pen Drive

The following is a description of the procedures to save the system calibration data stored in the removable hard disk to a USB Pen drive.

Step 1. Exit the measurement window.

System > Service Menu > Administorator Menu > Exit E5052B > OK

Step 2. Insert a USB Pen drive into the USB port in front panel.

Step 3. Open D:\Recovery drive.

Step 4. Click .D:\syscal from the explorer.

Step 5. Select Edit - Copy from menu bar of the explorer.

Step 6. Click the USB Pen drive of the explorer.

Step 7. Select Edit - Paste from menu bar of the explorer.

Step 8. Click .D:\limit from the explorer.

Step 9. Select Edit - Copy from menu bar of the explorer.

Step 10. Click the USB Pen drive of the explorer.

Step 11. Select Edit - Paste from menu bar of the explorer.

Step 12. Turn off the E5052B, and then remove the cord from the outlet before removing the Removable Hard Disk.