Calibration of the Touch Screen

Option Installation

Procedure

Calibration of the Touch Screen

When you have replaced the HDD on an E5052B, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen.

Step 1.

Press

.

Step 2. Press Service Menu.

Step 3.

Press Test Menu.

Step 4.

Press Adjust Touch Screen.

 

The touch screen calibration screen (Figure 2-4) appears.

Figure 2-4

Touch Panel Calibration Screen

 

Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the lower

 

 

left, upper right, and lower right. Touch the x marks in that order with your finger.

 

 

Touching the four locations described above with your finger automatically concludes the

 

 

touch screen calibration.

 

 

 

NOTE

 

With no operation on the touch screen calibration screen for a preset time, it automatically

 

 

closes and the previous measurement screen reappears.