Option Installation

 

Procedure

 

Recalling System Calibration Data from USB Pen drive to Removable

 

Hard Disk

 

The following is a description of the procedures to copy the system calibration data stored

 

in the USB Pen drive to the removable hard disk.

Step 1.

Exit the measurement window.

 

System > Service Menu > Administorator Menu > Exit E5052B > OK

Step 2.

Insert a USB pen drive in to the USB port in front panel.

Step 3.

Open D:\Recovery drive.

Step 4.

Click syscal folder in the USB Pen drive from the explorer.

Step 5.

Select Edit - Copy from menu bar of the explorer.

Step 6.

Click the D:\ of the explorer.

Step 7.

Select Edit - Paste from menu bar of the explorer.

Step 8.

Click limit folder in the USB Pen drive from the explorer.

Step 9.

Select Edit - Copy from menu bar of the explorer.

Step 10.

Click the D:\ of the explorer.

Step 11.

Select Edit - Paste from menu bar of the explorer.

Step 12.

Restart E5052B