Servicing the Interface Module
Removing and Replacing the HP 16500H Interface Module
6Install the new interface module by reversing this procedure.
7Test the performance of the interface board.
Refer to "Verifying the interface module performance" in this chapter for the performance verification test procedure.
8Reconnect the HP 16505A or HP 16501A to the interface board, or reconnect the interface board to your network.
If you are reconnecting to a network, refer to the LAN Administartor’s and Service Guide, or check with your system administrator for reconnecting instructions.
The Ether Address of the replacement interface module will be different than the other module. At powerup, the HP 16500B operating system reads the address of the interface module, then records the address in memory.
If your network configuration did not change, the addresses and information in the configuration menus will remain the same.