Servicing the Interface Module

Verify the Interface Module Performance

Procedure

This procedure verifies the performance of the interface module. To verify

 

performance of the HP 16500B Logic Analysis System or the optional

 

measurement modules, refer to the Service Guides for those products.

1If the screen appears blank, the interface module or cable may be defective.

a Remove power from the HP 16500B, disconnect the power cable, then remove the top cover.

b Remove and inspect the interface module cable for signs of damage. If the cable appears defective, replace it.

c Ensure that the cable is connected according to the installation procedures and that all cables are properly seated. For installation procedures, refer to chapter 2, "Installing the Interface Module."

d Install the top cover of the HP 16500B mainframe, connect the power cable, and reapply power.

2Touch the blue field in the upper-left corner of the HP 16500B screen. Select System from the pop-up menu .

3Touch the blue field immediately to the right. In the pop-up menu, select Configuration.

4Verify that "Ethernet" is available in the Communications menu on the HP 16500B System Configuration screen.

Ethernet should be listed here

3-3