Service Information

4

Separate the carrier and measurement board.

There are 32 relays as shown.

Turn to the opposite side of the measurement board and ensure that only the defective relays are desoldered.

Upon replacing the relay(s), reset the relay cycle count to zero by issuing the following SCPI command:

DIAGnostic:RELay:CYCLes:CLEar (@<ch_list>)