SECTION 5.0: TOUCHSCREEN DISPLAYS

The 6870 Freezer/Mill is programmed and operated through a series of touchscreen displays. Transition between screens, and all programming and operating commands, are done by touching the screen with a stylus. Do not use a sharp point as this can damage or deface the screen.

5.1 LOGO SCREEN

When you switch on the power, the LOGO screen appears first. This says “SPEX SamplePrep 6870 Freezer/Mill” and includes a graphic representation of the Freezer/Mill mascot. Touch the LOGO screen anywhere to transfer to the CURRENT SETTINGS screen.

5.2 CURRENT SETTINGS SCREEN

This screen displays whatever program is currently set up to run. Here is a typical program, as displayed:

CYCLES

3

PRECOOL

10 MIN

RUN TIME

2 MIN

COOL TIME

2 MIN

RATE

10 CPS

SETTINGS

Start SCRN

CURRENT SETTINGS SCREEN

If the Freezer/Mill has a loaded vial in the coil, is filled with LN, and the lid is down and latched, this is what happens when the mill is started with these settings:

1)Precool: the coil is inactive while the vial cools under LN for 10 minutes.

2)First cycle: the coil is activated and the impactor grinds for 2 minutes at a rate of 10 cycles per second (20 impacts per second).

3)Cool time: the coil is inactive for 2 minutes while it and the sample cool back down.

4)Second cycle: the coil is activated and the sample is ground for another 2 minutes.

5)Cool time: the coil is inactive for 2 minutes.