SECTION 8.0: TROUBLE-SHOOTING GUIDE, PAGE 1 OF 2

 

SECTION 8.0: TROUBLE-SHOOTING GUIDE, PAGE 1 OF 2

 

PROBLEM

CAUSE

SOLUTION

1

The unit does not turn on

No power

Make sure power cord is plugged into

 

 

 

outlet and inlet

2

The unit does not turn on

No power

Check outlet for power and correct as

 

 

 

required

3

The unit does not turn on

Power switch not turned

Switch unit on

 

 

on

 

4

The unit does not turn on

Blown fuse

Replace fuses

5

The unit does not turn on

Blows fuses repeatedly

Electrical short (RTF)*

6

Status Screen displays

Safety interlock switch

Close the lid and latch

 

“STATUS: LID UP”

not engaged

 

7

Status Screen displays

Lid is not closed

Push the lid closed and latch into place

 

“STATUS: LID UP”

completely

 

8

Status Screen displays

Liquid nitrogen sensor

Pour more liquid nitrogen into the tub

 

“STATUS: LOW LN LEVEL” )

senses inadequate liquid

 

 

 

nitrogen level

 

9

Coil gets very warm

Low liquid nitrogen

 

 

levels

10

Impactor doesn’t shuttle back

Water in vial has frozen

 

and forth

impactor

11

Impactor doesn’t shuttle back

Impactor is magnetized

 

and forth

 

12

Impactor doesn’t shuttle back

Too much sample is in

 

and forth

the vial or sample pieces

 

 

are too large

13

Flanged End Plug doesn’t come

End plug/tube joint too

 

off cold vial.

tight when very cold.

*RTF=Return to Factory for repair

Pour more liquid nitrogen into the tub

Remove vial and replace vial and contents with a dry unit or RAP ends of the vial on counter to free impactor

Remove impactor, turn end-for-end, or demagnetize impactor and re-insert

Remove some sample from the vial and start again

Allow vial to warm up