| SECTION 8.0: | ||
| PROBLEM | CAUSE | SOLUTION |
1 | The unit does not turn on | No power | Make sure power cord is plugged into |
|
|
| outlet and inlet |
2 | The unit does not turn on | No power | Check outlet for power and correct as |
|
|
| required |
3 | The unit does not turn on | Power switch not turned | Switch unit on |
|
| on |
|
4 | The unit does not turn on | Blown fuse | Replace fuses |
5 | The unit does not turn on | Blows fuses repeatedly | Electrical short (RTF)* |
6 | Status Screen displays | Safety interlock switch | Close the lid and latch |
| “STATUS: LID UP” | not engaged |
|
7 | Status Screen displays | Lid is not closed | Push the lid closed and latch into place |
| “STATUS: LID UP” | completely |
|
8 | Status Screen displays | Liquid nitrogen sensor | Pour more liquid nitrogen into the tub |
| “STATUS: LOW LN LEVEL” ) | senses inadequate liquid |
|
|
| nitrogen level |
|
9 | Coil gets very warm | Low liquid nitrogen |
|
| levels |
10 | Impactor doesn’t shuttle back | Water in vial has frozen |
| and forth | impactor |
11 | Impactor doesn’t shuttle back | Impactor is magnetized |
| and forth |
|
12 | Impactor doesn’t shuttle back | Too much sample is in |
| and forth | the vial or sample pieces |
|
| are too large |
13 | Flanged End Plug doesn’t come | End plug/tube joint too |
| off cold vial. | tight when very cold. |
*RTF=Return to Factory for repair
Pour more liquid nitrogen into the tub
Remove vial and replace vial and contents with a dry unit or RAP ends of the vial on counter to free impactor
Remove impactor, turn
Remove some sample from the vial and start again
Allow vial to warm up
Page 24 of 25 | SPEX SamplePrep LLC | 9.24.07 |