Optional Tap
Points
For
Digital TC
Total Control Acid In
Tap Point
Locations
Water
Out
Always Insert
tubing
into appropriate
inlet/outlet.
See Digital TC
manual
for drilling instructions.

INSTALLATION

Pool Chemistry Controller Flow Cell Plumbing

Optional Tap Points

 

 

 

For

 

 

 

 

Digital TC

 

 

It is necessary to have a constant flow to the flow cell to ensure accurate water

Total Control

 

 

 

 

 

 

sampling. This is done by locating the water sample tubes as indicated in the

 

 

 

 

diagram on pg. 16.

Acid In

 

 

Have these tools on hand prior to starting this step:

Tap Point

 

Water Out

 

 

 

Locations

 

 

Drill

 

 

 

 

• 1/2” standard drill bit

 

 

 

 

Utility knife

Always Insert tubing

 

 

 

 

 

 

 

 

into appropriate

 

 

Water Sample Feed Tube

inlet/outlet.

 

 

for drilling instructions.

 

 

 

 

See Digital TC manual

 

 

A hole must be drilled into the manifold assembly for installation of the injector fitting for the Water Sample Feed Line.

1.Turn the pump off, disconnect the cell and tri-sensor cords, and remove the manifold piping (without the cell) to a clean working area. See diagram above for which unions to remove (marked in diagram).

2.Locate the label on the manifold assembly marked “Optional Tap Points” (marked in diagram), and drill the

appropriate mounting hole adjacent to where indicated on the label marked “Water Out”. Do not drill through the label! Use a drill stop attachment if available, or wrap a piece of masking tape, exposing 1” of the tip

of the drill bit as a stop indicator. This will prevent drilling through the far side of the PVC, damaging the manifold.

Note: Use care to drill the hole perpendicular and deburr, for a leak-free connection of the injector fitting.

Do not over drill or ream the hole as oversized holes will leak and are not covered under warranty.

3. Insert the Uniseal fitting into the hole.

Note: Do not use any lubrication or sealing compounds!

4.Insert the Black Water sample tube approx 1” into the Uniseal fitting.

5.Roll out the tubing and connect into the Flow Cell Inlet valve (left side), ensuring a squared edge.

6.Leave enough slack in the tubing to secure the tubing with ty-wraps.

 

!Caution

 

Avoid any sharp bends and over tightening the ty-wrap to the tubing,

 

 

which can hinder proper flow of the chemicals

 

 

 

 

 

 

Water Sample Return Tube

The Flow cell sample return tube must be connected to the suction side of the circulation pump. This can be done by drilling a ½” hole in the pipe and using the Uniseal fitting, or by Drill and Tapping, then threading the Stenner ¼” NPT fitting in the pipe.

Using the Uniseal fitting (Turn Pump Off prior to drilling into the pipe):

1.Mark a spot on a straight section of suction pipe into the pump. Make sure it is after all suction valves and directly into the pump suction.

2.Drill a ½” hole. Use a drill stop attachment if available, or wrap a piece of masking tape, exposing 1” of the tip of the

drill bit as a stop indicator. This will prevent drilling through the far side of the PVC, damaging the pipe. Note: Use care to drill the hole perpendicular and deburr, for a leak-free connection of the injector fitting. Do not over drill or ream the hole as oversized holes will leak and are not covered under warranty.

3.Insert the Uniseal Fitting into the hole.

4.Insert the black water sample tubing into the suction side fitting.

5.Roll out the tubing and connect into the Flow Cell Outlet valve (right side), ensuring a squared edge.

6.Leave enough slack in the tubing to secure the tubing with ty-wraps.

7.Install the ORP and pH sensors as shown on pg. 19.

8.Turn the pump on and check for leaks.

A 6” section of tubing can be inserted into the water sample valve (middle valve) to allow better access to take a water sample from the flow cell.

21