28 MMD 540H Hybrid Multimeter Electrical Specifications
DC Current
Range | Resolution | Accuracy |
|
|
|
40.00mA | 10uA | ±1.5% of rdg ±3 dgts |
|
|
|
400.0mA | 100uA |
|
|
|
|
20A | 10mA | ±2.5% of rdg ±5 dgts |
|
|
|
Overload Protection: 0.5A / 250V and 20A / 250V Fuse.
Maximum Input: 400mA dc or 400mA ac rms on uA / mA ranges,20A dc or ac rms on 20A range.
AC Current
Range | Resolution | Accuracy |
|
|
|
40.00mA | 10uA | ±1.8% of rdg ±5 dgts |
|
|
|
400.0mA | 100uA |
|
|
|
|
20A | 10mA | ±3.0% of rdg ±7 dgts |
|
|
|
Overload Protection: 0.5A / 250V and 20A / 250V Fuse.
Frequency Range: 50 to 60 Hz
Maximum Input: 400mA dc or 400mA ac rms on uA / mA ranges, 20A dc or ac rms on 20A range.
Resistance
Range | Resolution | Accuracy |
|
|
|
400.0Ω | 0.1Ω | ±1.2% of rdg ±4 dgts |
|
|
|
4.000kΩ | 1Ω | ±1.0% of rdg ±2 dgts |
|
|
|
40.00kΩ | 10Ω | ±1.2% of rdg ±2 dgts |
|
|
|
400.0kΩ | 100Ω |
|
|
|
|
4.000MΩ | 1kΩ |
|
|
|
|
40.00MΩ | 10kΩ | ±2.0% of rdg ±3 dgts |
|
|
|
Capacitance
Range | Resolution | Accuracy |
|
|
|
4.000nF | 1pF | ±5.0% of rdg±50 dgts |
|
|
|
40.00nF | 10pF | ±5.0% of rdg ±7 dgts |
|
|
|
400.0nF | 0.1nF | ±3.0% of rdg ±5 dgts |
|
|
|
4.000uF | 1nF |
|
|
|
|
40.00uF | 10nF |
|
|
|
|
400.0uF | 0.1uF |
|
|
|
|
4.000mF | 0.001mF | ±10% of rdg ±10 dgts |
|
|
|
40.00mF | 10.00mF |
|
|
|
|
Robert Bosch LLC | MMD 540H Instruction Manual Version 1 February 2008 |