Braun BP2550 manual Specifications

Models: BP2550

1 62
Download 62 pages 37.84 Kb
Page 12
Image 12

Specifications

Model:

BP 2550

Operation principle:

Oscillometric method

Display:

Liquid Crystal Display

Measuring range:

0 – 300 mmHg (cuff pressure)

 

40 – 160 / min (pulse rate)

Blood pressure

 

measurement:

50 mmHg (minimum diastolic value)

 

250 mmHg (maximum systolic value)

Laboratory accuracy:

± 3 mmHg (cuff pressure)

 

± 5 % of reading (pulse rate)

Clinical accuracy:

According to AAMI-SP 10 with auscultatory reference:

 

< 5 mmHg systematic offset

 

< 8 mmHg standard deviation

Inflation:

Automatic, adapted to blood pressure

Deflation:

Electronic control valve adapted to heart frequency

Power supply:

2 batteries, type LR 03 (AAA)

 

(e.g. Duracell Ultra)

Operating temperature:

+ 10 °C to + 40 °C

Storage temperature:

– 20 °C to + 60 °C

Humidity:

Up to 85% RH

Cuff:

Suitable for wrist circumference of 13.5 to 20 cm

Clinical validation:

According to AAMI-SP10 with auscultatory reference.

 

(Please note that a validation against intra-arterial

 

measurements can lead to different results for devices

 

which are calibrated against auscultatory reference.)

12

Page 12
Image 12
Braun BP2550 manual Specifications