Replacing the batteries

The thermometer is supplied with 2 lithium cells, type CR 2032/ DL 2032. Insert new batteries when the battery symbol appears on the display.

Using the tip of a ball-point pen, press the battery door lock to open the battery compartment.

Using your finger, press the battery release.

Remove the batteries and replace with new batteries, making sure the poles are in the right direction.

Slide battery door back in until it snaps in place.

To protect the environment, dispose of empty batteries at your retail store or at appropriate collections sites according to national or local regulations.

To dispose of the thermometer, please return it to a Welch Allyn Service Centre.

Calibration

The thermometer is initially calibrated at the time of manufacture. If at any time you question the accuracy of measurement or the performance of the thermometer, contact a Welch Allyn Service Centre. To determine the date of manufacture, refer to the Lot number located in the battery compartment. The first digit refers to the last digit in the year of manufacture. The next two digits refer to the calendar week in the year of manufacture.

For example: LOT 842...

This thermometer was manufactured during the 42nd week in 1998.

Product specifications

Displayed temperature range: 20 - 42.2 ºC (68 - 108 ºF)

Operating ambient temperature range: 10 - 40 º C (50 - 104 ºF)

Display resolution: 0.1 ºC or ºF

Accuracy: 35.5 - 42.0 ºC (98.0 - 107.6 ºF): ± 0.2 ºC

outside this range: ± 0.3 ºC

Battery life: 5000 measurements

Type BF equipment / Subject to change without notice.

The Braun ThermoScan PRO3000 has been found to be immune to electromagnetic fields and does not produce any measurable

RF emissions.

 

This product conforms to the provisions of the

0297

EC directive 93/42/EEC (Medical Device Directive).