5.1.1.3 Sample Program for CR5000

'CR5000

'This example program measures a single 109 Thermistor probe 'once a second and stores the average temperature every 10 minutes.

'Declare the variable for the temperature. Public Air_Temp

'Declare variables for the raw measurement, thermistor resistance, and ln(resistance): Dim V_Vx, Rtherm, lnRt

'Define a data table for 10 minute averages: DataTable (AvgTemp,1,1000)

DataInterval (0,10,min,10)

Average (1,Air_Temp,IEEE4,0)

EndTable

BeginProg

Scan (1 ,sec,5,0)

'Measure the 109 probe. The result is V/Vx:

BrHalf (V_Vx,1,mV5000,3,Vx1,1,5000,True ,0,_60Hz,1.0,0) 'Calculate reistance:

RTherm=24900*(1/V_Vx-1)

'Calculate the natural log of the resistance: lnRt=Log(Rtherm)

'Apply the Steinhart and Hart equation and convert to degrees C in one step: Air_Temp=1/(1.129241e-3+2.341077e-4*lnRt+8.775468e-8*(lnRt^3))-273.15'Call the data table:

CallTable AvgTemp

NextScan

EndProg

5.2 Edlog

In Edlog, Instruction 5 is typically used to measure the 109SS resistance. Instruction 55 is used to apply the Steinhart and Hart equation. Instruction 55 does not allow entering the coefficients with scientific notation. In order to use this instruction with as much resolution as possible, the ln resistance term is pre scaled by 10-3. This allows the first order coefficient (B) to be multiplied by 103, and the 3rd order coefficient (C) to be multiplied by 109 (see Section 5.2.1).

5.2.1 Example Edlog Program

TABLE 5-2. Wiring for Example Program

Color

Description

CR10X

Black

Excitation

E1

Red

Signal

SE1

Purple

Signal Ground

AG

Clear

Shield

G

7