Model 109SS Temperature Probe
9: Z=X+F (P34) |
|
| |
1: | 8 | X Loc [ Tk | ] |
2: | F |
| |
3: | 9 | Z Loc [ Air_Temp ] | |
10: If time is (P92) |
|
| |
1: | 0 | Minutes (Seconds | |
2: | 10 | Interval (same units as above) | |
3: | 10 | Set Output Flag High (Flag 0) | |
11: Real Time (P77) |
| ||
1: | 110 | Day,Hour/Minute (midnight = 0000) | |
12: Average (P71) |
|
| |
1: | 1 | Reps |
|
2: | 9 | Loc [ Air_Temp ] | |
*Table 2 Program |
|
| |
02: | 0.0000 | Execution Interval (seconds) |
*Table 3 Subroutines
End Program
5.3 Electrical Noisy Environments
AC power lines, pumps, and motors, can be the source of electrical noise. If the 109SS probe or datalogger is located in an electrically noisy environment, the 109SS probe should be measured with the 60 or 50 Hz rejection option as shown in the examples in Section 5.1.1.2 and Section 5.2.1.
5.4 Long Lead Lengths
Additional settling time may be required for lead lengths longer than 300 feet, where settling time is the delay before the measurement is made.
For the
‘Therm109 ( Dest, Reps, SEChan, ExChan, SettlingTime, Integ, Mult, Offset )
Therm109(T109_C,1,1,1,20000,_60Hz,1.0,0.0)
In Edlog, use the DC Half Bridge instruction (P4) with a 20 millisecond delay as shown below. Use P4 in place of P5 in Section 5.2.1 (the instructions that follow P5 to convert the measurement result to temperature are still required).
9