Model 109SS Temperature Probe

Example Program for CR10X

;{CR10X}

 

 

 

;

 

 

 

 

*Table 1 Program

 

 

 

01:

1

Execution Interval (seconds)

1: AC Half Bridge (P5)

 

 

1:

1

Reps

 

 

2:

25

2500 mV 60 Hz Rejection Range

3:

1

SE Channel

 

 

4:

1

Excite all reps w/Exchan 1

5:

2500

mV Excitation

 

 

6:

1

Loc [ V_Vx

]

 

7:

1.0

Mult

 

 

8:

0.0

Offset

 

 

2: Z=1/X (P42)

 

 

 

1:

1

X Loc [ V_Vx

 

]

2:

2

Z Loc [ Vx_V

 

]

3: Z=X+F (P34)

 

 

 

1:

2

X Loc [ Vx_V

 

]

2:

-1

F

 

 

3:

3

Z Loc [ Vx_V_1

]

4: Z=X*F (P37)

 

 

 

1:

3

X Loc [ Vx_V_1

]

2:

24900

F

 

 

3:

4

Z Loc [ Rtherm

 

]

5: Z=LN(X) (P40)

 

 

 

1:

4

X Loc [ Rtherm

]

2:

5

Z Loc [ lnRt

]

 

6: Z=X*F (P37)

 

 

 

1:

5

X Loc [ lnRt

]

 

2:

.001

F

 

 

3:

6

Z Loc [ Scal_lnRt ]

7: Polynomial (P55)

 

 

1:

1

Reps

 

 

2:

6

X Loc [ Scal_lnRt ]

3:

7

F(X) Loc [ 1_Tk

]

4:

.001129

C0

 

 

5:

.234108

C1

 

 

6:

0.0

C2

 

 

7:

87.7547

C3

 

 

8:

0.0

C4

 

 

9:

0.0

C5

 

 

8: Z=1/X (P42)

 

 

 

1:

7

X Loc [ 1_Tk

 

]

2:

8

Z Loc [ Tk

]

 

 

 

 

 

 

8