Elo TouchSystems AMM15TK2-TEW operation manual Cautions

Models: AMM15TK2 AMM15TK2-TEW AMM15TK2-TE

1 32
Download 32 pages 38.27 Kb
Page 6
Image 6

Cautions

1.The AC Adapter must be plugged into a Grounded power outlet.

2.Use only the proprietary AMM15TK2 power supply for the AMM15TK2 monitor. Make a proper connection by ensuring that the shrink tubing completely secures the connection between the DC power cord and the extension cord.

3.Turn power off when unit is not in use.

4.Never operate the unit right after having transported from a cold location directly to a warm location.

5.Do not expose the monitor to moisture or directly apply liquid cleaners directly to the screen. Spray the cleaning solution into a soft cloth and clean gently.

6.Handle the monitor with care. Do not strike or scratch the screen.

7.Do not block the monitor cooling vents. The monitor is cooled by natural convection and has no fan.

8.Do not force the monitor past 28 degrees of vertical when adjusting the screen position. (For monitors equipped with stands only.)

9.Remove the power module and connection when transporting the unit.

10.Save the original carton and associated packing material. They will be useful should you have to transport or ship the unit.

11.Allow adequate air circulation to prevent internal heat buildup.

12.Do not place the unit on surfaces (rugs, blankets, etc.) or near materials (curtains, draperies) that may block the ventilation slots.

13.Do not install the unit near sunlight, excessive dust, mechanical vibration or shock.

14.The unit is designed for operation in a horizontal position. Never operate the unit in a vertical position.

15.Keep the unit away from equipment with strong magnets (i.e. a large loudspeaker.)

16.Do not expose the monitor to moisture or excessive dust.

17.Equipment with SIP/SOP connectors should either comply with IEC 60601-1 and/or IEC 60601-1-1 harmonized national standard or the combination should be evaluated. Do not touch the patient with signal input or output connectors.

18.Use only a hospital grade power supply cord.

19.This equipment generates, uses, and can radiate radio frequency energy. If not installed correctly and or not used in accordance with these instructions, it may cause harmful interference with other devices. This may be determined by turning the equipment off and on. The user is encouraged to try to correct the interference through one or more of the following measures:

•Reorient or relocate the receiving device.

•Increase the separation distance between the equipment.

•Connect the equipment to an outlet on a circuit different from that to which the other device(s) are connected.

•Consult the manufacturer or field service technician for help.

20.Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle labeled “Hospital Only” or “Hospital Grade.”

Note To connect to an international power supply, use a an attachment plug appropriate for the power outlet.

Note Refer to the “Electromagnetic Compatibility” (EMC) section of this manual to ensure EMC. The AMM15TK2 must be installed and operated according to the EMC information provided in this manual.

6

Page 6
Image 6
Elo TouchSystems AMM15TK2-TEW operation manual Cautions