Emerson F-SERIES SENSOR Orienting the Sensor, Recommended sensor orientations, Vertical pipeline

Models: F-SERIES SENSOR

1 26
Download 26 pages 16.83 Kb
Page 14
Image 14
Step 2 Orienting the Sensor

Orienting the Sensor

Step 2 Orienting the Sensor

The sensor will function properly in any orientation if the sensor tubes remain filled with process fluid. Micro Motion recommends orienting F-Series sensors as shown in Figure 8.

Figure 8 Recommended sensor orientations

 

 

 

 

 

 

Gases

 

 

 

 

 

 

 

 

 

 

Liquids

 

Slurries and

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

self-draining applications

 

 

 

 

 

 

Tubes up

 

 

 

 

 

 

 

 

 

 

Tubes down

 

Flag mount

 

Horizontal pipeline

 

 

 

 

 

Horizontal pipeline

 

Vertical pipeline

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow

Self-draining applications

The F-Series sensor is self-draining when it is installed in a vertical pipeline (see Figure 8). For CIP applications it is recommended to use the generally accepted flow velocity of at least 1.5 m/sec for cleaning the sensor.

Flow direction arrow

The sensor has a flow direction arrow (see Figures 1–5) to help you configure the transmitter for flow direction. If possible, install the sensor so that the flow direction arrow matches actual process flow.

Vertical pipeline

If the sensor is installed in a vertical pipeline, liquids and slurries should flow upward through the sensor. Gases may flow upward or downward.

12

Sensor Installation: F-Series

Page 14
Image 14
Emerson F-SERIES SENSOR Orienting the Sensor, Recommended sensor orientations, Self-drainingapplications