Micro Motion F-SeriesSensor
Installation Manual
P/N 20002298, Rev. C September
Page
Troubleshooting
Customer service
Before You Begin
European installations
Specifications
Installation options
Definitions
Calibration tag Core processor housing
Figure 1 F-Seriessensor with core processor
Process connection
Purge fitting Core processor housing
Sensor Installation F-Series
Core processor housing Calibration tag
Before You Begin
Model 1700/2700 transmitter
Sensor Installation F-Series
Figure 3 F-Seriessensor with junction box
Before You Begin
Approval tag Junction box Calibration tag
Before You Begin
Sensor Installation F-Series
Model IFT9701 transmitter Calibration tag
Approval tag Process connection
Before You Begin
Sensor Installation F-Series
Transmitter, core processor, or
junction box core processor shown
Step 1 Determining a Location
Hazardous area installations
Full flow tubes
Environmental limits
Determining a Location
Determining a Location
Table 1 UL and CSA ambient temperature limits
Sensor models
Electronics
Table 2 Maximum cable lengths
Maximum wiring distances
Pipe run
Valves
Figure 8 Recommended sensor orientations
Step 2 Orienting the Sensor
Self-drainingapplications
Flow direction arrow
Figure 9 Mounting an F-Seriessensor
Step 3 Mounting the Sensor
Mounting bracket instrument pole
Mounting bracket wall mount
Slide the electronics into the bracket groove
Step 4 Wiring
Core processor to a 4-wireremote transmitter
Installation options
Figure 12 Wrapping the shield drain wires
Figure 13 Applying the heat shrink
Figure 15 Gland body and core processor housing
Figure 14 Folding the cloth shield
Sensor Installation F-Series
4.Tighten the screws to hold the wires in place
Step 5 Grounding
Pressure Ratings at High Temperatures
Grounding
Purge Fittings
Case purging procedure
Removing a purge plug
Purge rate
Purge Fittings
Time1
Sensor model
New and unused equipment
Return Policy
Used equipment