Emerson IP258 manual 0.00, P150

Models: IP258

1 60
Download 60 pages 14.91 Kb
Page 21
Image 21
0.00

After a few secnds the display will show the following.

 

 

 

SET SPAN

0

 

 

 

 

 

 

In sample 1

0

Press ENT to set 33.6dB 44.2dB

The bottom line of the display shows the attenuation caused by the sludge. During a desludge these readings will fluctuate due to random variations in the sludge density. When the readings are reasonably stable press ENT at the same time that a sample is taken. This will store the reading from the sensor.

 

SET

SPAN

0

 

Span1 is

now set

0

ESC to finish

for Span2

Press the down arrow key to access the sample 2 and sample 3 screens (if required).

Repeat the above procedure for samples 2 and 3. The span is now set and ready for input of laboratory result (see later). When the sample procedure is complete press ESC until the normal display shows.

For best accuracy the three samples should be taken over a period of approximately 2 to 3 minutes. The samples should now sent for laboratory analysis to establish the actual suspended solids content.

When the laboratory analysis results are available, the span calibration can be completed by entering the solids content (in % Solids) into “Lab Value 1” (P150).

If required, an estimated value can be used until the laboratory analysis results are available.

Note:

Each time SPAN 1 is set the other two SPAN and LAB values are cleared and are not used

 

until new Span2 and Span3 calibrations are carried out. The MSM400 calculates the

 

average of all the valid span calibrations (a valid span calibration is one which has both a

 

span ‘N’ and lab val ‘N’ value). The calibrations must be carried out in numerical order

 

i.e. Span 1, then Span 2 and then Span 3. These values should be carried over a short

 

period of time i.e. 2 to 3 minutes.

6.5AUTO CAL LAB VALUES

To complete the AUTO CAL the samples must now be analysed and the results entered into the lab value menu. Access the LAB VALS menu by highlighting this option in the AUTOCAL menu and pressing ENT.

 

LAB VAL

0

 

LabVal 1

0

LabVal 2

LabVal 3

Highlight the number of the LabVal to be entered and press ENT

 

LAB

VAL

0

 

LabVal 1

%

0

0.00

P150

 

Highlight the digit to be edited by pressing left or right arrows. To change the LabVal use the up and down arrows until the required value is showing, then use left or right arrow to move on to next digit. The units of this parameter are always % suspended solids. When the correct value is displayed press ENT to store the value. Repeat this step to input LabVal 2 and LabVal 3 if required (please note that better accuracy is achieved if all three samples and corresponding LabVals are entered). Press ESC until the normal display shows.

It is important to note that until the LabVal (1,2 or 3) are entered the system will use the default value of attenuation for the calibration.

IP258

21

Page 21
Image 21
Emerson IP258 manual 0.00, P150