D.6.0 How to drive a Psion based SMART Communicator

D.6.0 How to drive a Psion based SMART Communicator

Familiarity with the Psion

The Psion is supplied with several manuals to described its function. The main keyboard functions that are important are the yellow keys. Press “ON” and see the functions of the arrows to move the cursor around the selections in the menu. Press “EXE” to see the Psion functions. Note that a short cut to selecting a menu item is to press the key with the initial letter of the required selection – for example “O” switches the unit “OFF”.

Note that if the “CALC” option is selected, the keyboard changes from the “Alphabet” marked on the keys, to the “numbers” and symbols marked above the keys. The Psion selects the expected function required of the keyboard. This function can be changed back by depressing the “SHIFT” key.

MOBREY Menu

If “MOBREY” is selected from the start up menu on the Psion, by pressing “EXE” the space available in the A memory of the Psion is checked. If there is insufficient then a message is displayed and the program returns to the main menu. Data, Note Pad or Diary files in A must be removed to allow the program to run. The SMART Communicator then establishes whether the SMART interface unit is plugged in or not.

10:41 MOBREY Find Save Diary Calc Time Notes World Alarm

When the SMART interface is not connected, the Communicator does not look for an instrument – it offers OFF-LINE programming.

When the SMART interface is connected to the Psion, on selecting the “MOBREY” menu item the programme seeks an instrument that is expected to be functioning on a 4-20mA loop (the loop that is expected to be connected to the SMART interface cables).

MOBREY V3.1

SEEKING INSTRUMENT

PLEASE WAIT

If the loop is not powered, or the interface cables are not properly connected to the loop, or the loop impedance/resistance values are incorrect, the unit will fail to find the control unit.

NO SINGLE LOOP

INSTRUMENT CONNECTED

RETRY (Y)es or (N)o

Note that if the MSM400 control unit is programmed to respond as a numbered sensor on a multidrop loop, it will also fail to communicate to the SMART communicator at this point.

IP258

46

Page 46
Image 46
Emerson IP258 manual MOBREY Find Save Diary Calc Time Notes World Alarm, Mobrey Seeking Instrument Please Wait