5.2 I/O Module Controls 5.2.2 DIO Module(s) – Output configuration

·Go to the line you want with the -key or the -key. If there are more than eight source modules available, you have to press the F5 key to go to another menu page.

·Select the tag of the module with the 8 -key or the -key:

The display will go back to the menu "DIO-Module Outputs" automatically. The tag of the module selected will appear in the line "Signal comes from".

3)Select any signal:

·Change to the line "Choose signal..." with the -key.

·Press the 8 -key or the -key to change to the submenu "Signals":

TAG

95.00 ppm

-- Signals --

Function control

Maintenance request

Failure

Cal. in progress

Zero in progress

Span in progress

Zero failed

Span failed

Measure

 

<<<

Back...

>>>

F1

F2

F3

F4

F5

·Go to the line you want with the -key or the -key. With the F5 key you can go to further menu pages containing also available signals:

Range overflow

Range underflow

Flow too low

Flow too high

SYS:Valve-1, -2, ... , -32

and further signals, if need be:

All signals of the control module and of the analyzer modules joint to the platform or MLT analyzer (e.g. CLD, FID, MLT AM) are available in the menu "Signals".

·Select the signal with the 8 -key or the -key:

The display will go back to the menu "DIO-Module Outputs" automatically. The signal selected will appear in the line "Signal name".