5.2 I/O Module Controls 5.2.2 DIO Module(s) – Input configuration

·Select the tag of the module with the 8 -key or the -key:

The display will go back to the menu "DIO-Module Inputs" automatically. The tag of the module selected will appear in the line "Signal comes from".

MLT25/CH2/R2

95.00 ppm

-- Analyzer Modules --

Control Module

MLT25/CH1

MLT25/CH2

MLT25/CH3

MLT25/CH4

MLT25/CH5

Measure

 

<<<

Back...

>>>

F1

F2

F3

F4

F5

3)Select any function:

·Change to the line "Choose function..." with the -key.

·Press the 8 -key or the -key to change to the submenu "Functions":

TAG

95.00 ppm

-- Functions --

AM:Zero-Cal

AM:Span-Cal

AM:Range-1

AM:Range-2

AM:Range-3

AM:Range-4

SYS:Zero-Cal

SYS:Zero/Span-Cal

Measure

 

<<<

Back...

>>>

F1

F2

F3

F4

F5

·Go to the line you want with the -key or the -key. With the F5 key you can go to further menu pages containing also available signals:

SYS:Program-Cal

SYS:Cancel-Cal

SYS:CAL-Test-Mode

SYS:AM-Zero-Gas

SYS:AM-Span-Gas1, 2, 3, 4

External function control

AM: Hold Outputs

AM-closed-valves