If the deviation of the two values is less then the allowed maximum calibration deviation (see menu "Tolerances", p.
The following figure shows the procedure of the stability controlled zero and span gas calibration:
Concentration
Start Span
Calibration
Averaging
over Ti
Ti
Start Zero
Calibration
Purge | Stability | Stability | Stability | Stability | Time |
Time | Time | Time | Time | Time |
|
Note:
The stability time and the averaging time are factory settings and cannot be changed by the customer. If you want to use other values, ask your service !
Line of variables "Analog output during calibration":
Here, you can determine if you want to hold the analog output and the limits of a local SIO in the MLT analyzer module during calibration. The
♦Tracking: The analog output signal is following to the signal during the whole calibra- tion procedure.
♦Holding: During the whole calibration procedure the analog output signal is fixed to the last signal before calibration. This adjustment may be used for instance to control limits with the analog output.
Sample Gas | Zeroing | Spanning | Sample Gas |
Analog Output |
|
|
|
Tracking |
|
| xV |
0V |
|
| |
Analog Output |
|
|
|
Holding |
|
|
|
5 - 10 | NGA 2000 |