5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Calibration procedure setup
Main Menu — Analyzer and
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Analyzer module setup |
|
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Calibration parameters |
|
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Calibration procedure setup |
|
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| TAG |
|
|
|
|
|
|
| 95.00 ppm |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
| Purge time: |
|
|
|
|
|
|
| 10 | s |
|
|
|
|
|
|
| ||||
| Procedure |
|
|
| 120 | s |
|
|
|
|
|
|
| ||||||||
| Analog output during calibration: |
|
|
|
| Tracking |
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
| |||||||||||
| Span ranges: |
|
|
|
|
|
| Together |
|
|
|
|
|
|
| ||||||
| Valve position: |
|
|
|
|
|
| Samplegas |
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
| Stability time: |
|
|
|
|
|
| 30 | s |
|
|
|
|
|
|
| |||||
| Averaging time: |
|
|
|
|
|
| 5 | s |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Measure |
|
|
| Channel |
| Back... |
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| F1 | F2 |
| F3 |
| F4 | F5 |
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the menu "Calibration Procedure Setup" you can set up the parameters for the calibration procedure of the zeroing and spanning in one channel. If you want to set up the parameters of the other channels of an MLT analyzer or analyzer module, press the F3 key.
Set up parameters:
¨Select any line of variables with the ↓
¨Select the variable with the 8
¨Select any digit with the ←
resp. select the whole parameter with the −
¨Confirm the new value with the 8
cancel and go back to the last value with the F2 key.
Lines of variables "Purge time" / "Procedure
"Stability time" / "Averaging time" – Informations about the calibration procedures setup:
The measurement signal should be stable in a certain range before zeroing and spanning. Therefore the analyzer cell should be purged enough with zero gas or span gas before calibration. Then, after the purge time, the stability time is running. During the stability time the average over two signals will be calculated. The averaging time determines the time difference between these two signals.
90003482(2)
NGA 2000 | 5 - 9 |