5.1 Analyzer Module Setup

5.1.1 Calibration Parameters – Tolerances

Main Menu — Analyzer and I/O-module expert configuration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer module setup

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calibration parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Tolerances

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

 

 

95.00 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

-- Tolerances --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Max. zero calibration deviation:

 

 

 

20.00 %

 

 

 

 

 

 

 

 

Max. span calibration deviation:

 

 

 

20.00 %

 

 

 

 

 

 

 

 

Check deviation:

 

 

 

 

 

 

 

Disabled

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Stability tolerance - range 1:

 

 

 

100.00 %

 

 

 

 

 

 

 

 

Stability tolerance - range 2:

 

 

 

100.00 %

 

 

 

 

 

 

 

 

Stability tolerance - range 3:

 

 

 

100.00 %

 

 

 

 

 

 

 

 

Stability tolerance - range 4:

 

 

 

100.00 %

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Last zero calibration:

 

 

 

 

 

 

 

Success

 

 

 

 

 

 

 

 

Last span calibration:

 

 

 

 

Success

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

 

 

 

Channel

 

Back...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

 

F3

 

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Tolerances" you can set up the parameters of the stability controlled calibration procedure (see p. 5-9/10) in one channel. Besides, you can disable the stability control and the maximum calibration deviation (deviation control ® deviation tolerance).

If you want to set up the parameters of the other channels of an MLT analyzer or analyzer module, press the F3 key.

Set up parameters:

¨Select any line of variables with the -key or the -key.

¨Select the variable with the 8 -key or the -key.

¨Select any digit with the -key or the -key and adjust a new value with the -key or the -key

resp. select the whole parameter with the -key or the -key.

¨Confirm the new value with the 8 -key or

cancel and go back to the last value with the F2 key.

Lines "Last zero calibration/Last span calibration":

The calibration status of the last calibration will be displayed in these two lines.